2) For test weights calibration mode, the following method should be used:
a) The Equivalent Load (Kg/m) =
DengthWeighSpanL
KgsTestWeight
Test Weights = Static weight applied on the weigh idlers
Weigh Span Length = Idler spacing X Number of weigh idlers
For example:
K g = 100
D = 2.4m
Kg/m = 41.67 kg/m
b) Simulated Constant For Test Weights can be calculated using the following formula:
The Simulated C onst ant =
1000
)()/(
mationforTestDurBeltLength mkg
×
For example:
The Belt Length for
Test
Duration = 180m
The Simulated Constant =
1000
180 67 .41 ×
= 7.5t
3) For test chains calibration mode, the following method should be used:
a) The Equivalent Load (Kg/m) is specified at the tail of the test chains
b) Simulated Constant For
Test
Chains can be calculated using the following formula:
The Simulated C onst ant =
1000
)()/(
mationforTestDurBeltLength mkg
×
For example:
The Belt Length for
Test
Duration = 180m
Kg/m = 30
The Simulated Constant =
1000
180 30 ×
= 5.4t
Note:
Note:
Note:
Note: The system setup is intended to program some basic parameters which determine whether
correct result can be achieved from the belt scale. These parameters include scale number and scale
capacity etc.
The following is an explanation to a part of these parameters.
[Zero Tracking]: Zero may drift due to various influence factors. When Zero Tracking is set, automatic
zeroing will be implemented in case of a total lower than the preset value.
[Zero Dead]: During a run period, a total lower than the preset Zero Dead will not be added into the
master total.
[Accumulated Pulses]: When the total is up to the preset Accumulated Pulses, an accumulated pulse
signal will be emitted.
[Pulse Width]: This option defines the ratio of high voltage level [1] and low voltage level [0] of pulse
width
[Control Mode]: There are two control modes selectable: automatic and manual.
[Set Mode ] : When automatic control mode is selected, two set modes are available for selection: local
and remote.