Revised 5/19/2010 6.3
Error Codes and Messages
Error
Code
Title
Message
Result, or Action
Required
Error Type Error Name
13
15
16
17
19
HEATER TEMPERATURE PROBLEM
The [reservoir / copper tube 1 / copper tube 2 / flex tube /
valve/nozzle] has dropped below acceptable operating
temperature. Base molds will be cleared and no more
cassettes will be embedded. The paraffin system heaters
will be turned off, and the instrument MUST be serviced
before embedding can continue.
Perform action/s
requested.
19
HEATER TEMPERATURE PROBLEM
The frame heater -<temperature> is out of acceptable
operating range <Min.-Max.>, instrument will now shut
down, empty input door and all output doors. Contact
Sakura Service Department.
14
18
20
HEATER TEMPERATURE PROBLEM
The [secondary reservoir / pusher pins / stager] [is / are]
outside acceptable operating temperature. Current
magazine will be finished and base molds cleared. The
instrument SHOULD be serviced before embedding
continues.
Perform action/s
requested.
21
22
HEATER TEMPERATURE PROBLEM
Heater [A / B] in input door chamber (#) has dropped
below acceptable operating temperature. The system will
not use this input chamber until temperature is within
normal range. The instrument MUST be serviced before
chamber (#) can be used.
Perform action/s
requested.
Heater [A / B] in input door chamber (#) has risen above
acceptable operating temperature. Cassettes in the
magazine will be embedded and base molds cleared. The
instrument SHOULD then be serviced before embedding
continues
23
HEATER THERMAL FUSE PROBLEM
During a check of the system heaters [reservoir /
secondary reservoir / copper tube 1 / copper tube 2 / flex
tube / valve/nozzle/ drip tray/ ID-(#)A/ ID-(#)B] thermal
fuse was discovered to be functioning improperly.
The instrument will not run until this problem is fixed.
Please contact maintenance for further assistance.
Perform action/s
requested.
24
25
HEATER TEMPERATURE PROBLEM
The [tamper / gripper] heater -<temperature> is out of
acceptable operating range <Min.-Max.>, base molds will
be cleared, no more cassette will be embedded.
Perform action/s
requested.
26
HEATER TEMPERATURE PROBLEM
The drip tray heater -<temperature> is out of acceptable
operating range <Min.-Max.>, instrument will continue
embedding as normal. Contact Sakura Service
Department.
Perform action/s
requested.
27
REQUEST
OUTPUT DOOR #<1–4> IS
NOT PRESENT
Please replace Output Door #<1–4>, then touch OK to
resume instrument operation.
Touch OK
EMPTY AND REPLACE
OUTPUT DOOR #1
Operation WILL NOT continue until Output Door #1 is in
place and empty. Once the Output Door has been
replaced, touch OK to resume instrument operation.
Touch OK
27
REQUEST
EMPTY AND REPLACE
OUTPUT DOOR #2
Operation WILL NOT continue until Output Door #2 is in
place and empty. Once the Output Door has been
replaced, touch OK to resume operation.
Touch OK
EMPTY AND REPLACE
OUTPUT DOOR #3
Operation WILL NOT continue until Output Door #3 is in
place and empty. Once the Output Door has been
replaced, touch OK to resume operation.
Touch OK
EMPTY AND REPLACE
OUTPUT DOOR #4
Operation WILL NOT continue until Output Door #4 is in
place and empty. Once the Output Door has been
replaced , touch OK to resume operation.
Touch OK