11
<Ex.1>When O
2
/N
2
Cylinder A is emptied:
Operate with O
2
/N
2
Cylinder A
O
2
/N
2
lamp is lit
When O
2
/N
2
Cylinder A in current use
is emptied, operation is switched from
Cylinder A to Cylinder B. At the time
the display shows internal O
2
level and
“A” alternately.
O
2
/N
2
lamp is blinking
If O
2
/N
2
Cylinder B is also emptied,
operation is switched from Cylinder A
to Cylinder B. The display shows
internal O
2
level and “A” alternately.
(Both Cylinder A and B are emptied)
O
2
/N
2
lamp is lit
<Ex.2>When O
2
/N
2
Cylinder B is emptied:
Operate with O
2
/N
2
Cylinder B
O
2
/N
2
lamp is blinking
When O
2
/N
2
Cylinder B in current use
is emptied, operation is switched from
Cylinder B to Cylinder A. At the time
the display shows internal O
2
level and
“b” alternately.
O
2
/N
2
lamp is lit
If O
2
/N
2
Cylinder A is also emptied,
operation is switched from Cylinder B
to Cylinder A. The display shows
internal O
2
level and “A” alternately.
(Cylinder A has priority to be displayed
when both Cylinder A and B are
emptied)
O
2
/N
2
lamp is blinking
A / B : In CO
2
setting mode, press the key to switch the CO
2
cylinder in use to another one.
(Ex. When the CO
2
cylinder A is currently used, press the key to switch to the CO
2
cylinder B. Press again to switch to the CO
2
cylinder A.)
In O
2
setting mode, press the key to switch the O
2
cylinder in use to another one.
(Ex. When the O
2
cylinder A is currently used, press the key to switch to the O
2
cylinder B. Press again to switch to the O
2
cylinder A.)