23
<SV and buzzer recovery time>
SV
Buzzer recovery time after buzzer is
cancelled by pressing BZ key
000 Buzzer does not recover
010 Approx. 10min
020 Approx. 20min
030 Approx. 30min
040 Approx. 40min
050 Approx. 50min
060 Approx. 60min
*Optional mode: Remote alarm can be linked with buzzer in optional function mode.
(The function code is not opened to users)
Remote alarm action
Buzzer solely activates
(F24=000)
Linked with buzzer
(F24=001)
Buzzer setting Initial: auto recovery
(F25=030)
Buzzer: auto recovery
Remote alarm: activating
Buzzer: auto recovery
Remote alarm: auto recovery
Inoperative auto
recovery (F25=000)
Buzzer: does not auto recover
Remote alarm: activating
Buzzer: does not auto recover
Remote alarm: does not auto
recover
Setting Remote alarm action
000 (initial) Not linked with buzzer
001 Linked with buzzer
16. Temperature offset value
The difference between temperature sensed in temperature control sensor and temperature in
internal 1/2H should be offset:
Offset value: +0.8°C
17. Storing in non-volatile memory
2 times comparison method: When the main power is supplied, a data storing in non-volatile
memory is called for 2 times to compare. If the data 1
st
called coincides with the data of 2
nd
called, the data would be stored in RAM. Otherwise, the comparison would be repeated until
the one coincides with another.
18. Example for displays
Current temperature +36.2 CO
2
level +5.0%
Set temperature +37.0 CO
2
level zero adjustment value –1.3%
Function F03 CO
2
gas error E01
(Service code input) UV lamp additional time
Bottom heater ratio 004
Lock Mode L 0

Other manuals for Sanyo MCO-5M