86
30
ALARMS & SAFETY FUNCTIONS
Alarm & Safety Situation Indication Buzzer Safety operation
If the ambient temperature sensor is
disconnected.
E09 is displayed alternately
with the temperature on the
temperature indicator.
Ambient
temperature sensor
abnormality
If the ambient temperature sensor is
short circuited.
E10 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone Remote alarm
CO
2
sensor
abnormality
If the output voltage of the CO
2
sensor
is abnormal.
E11 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone
CO
2
valve closes.
Remote alarm
O
2
sensor
abnormality
If the output voltage of the O
2
sensor
is abnormal.
E19 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone
N
2
/O
2
valve closes.
Remote alarm
Main heater
abnormality
If the upper limit alarm temperature
alarm operates, or if the main heater is
open circuit, or the main heater relay
short circuit.
E12 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone Remote alarm
Bottom heater
abnormality
If the bottom heater goes open circuit,
or the bottom heater relay short circuit.
E13 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone Remote alarm
Door heater
abnormality
If the door heater goes open circuit,
or the door heater relay short circuit.
E14 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone Remote alarm
Disconnection of
S.S.R for each
heater
If the relay of main heater, bottom
heater or sensor box heater goes open
circuit.
E16 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone Remote alarm
Low humidifying
water
If the water in the pan is about 0.8 liter. RH PAN lamp blinks. ----- -----
UV lamp failure When the UV lamp is burned out.
E18 is displayed alternately
with the temperature on the
temperature indicator.
Intermittent tone Remote alarm
Recommendation of
new UV lamp
The accumulated ON time is over
about
1,000 hrs.
UV indicator blinks when
UV lamp is OFF.
----- -----
• The alarm can be canceled by pressing the BUZZER key, but the remote alarm cannot be silenced. And the upper
limit temperature alarm cannot be silenced with the BUZZER key.
• E01 or E02 is cleared automatically when the gas is connected correctly and the buzzer is silenced with the
BUZZER key. Press BUZZER key to silence the alarm after changeover of gas supply line.
• If one of E05 to E17 is displayed, consult with a Sanyo sales representative or agent.
Operation after power failure
The set value is memorized by nonvolatile memory. Accordingly, the incubator resumes the operation
with setting before power failure.