EasyManua.ls Logo

Sanyo MCO-5M - Performance

Sanyo MCO-5M
48 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
100
44
PERFORMANCE
Temperature control range
Ambient temperature +5
o
C to 50
o
C (ambient temperature; 5
o
C to 35
o
C)
Temperature distribution
± 0.25
o
C* (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
Temperature variation
± 0.1
o
C (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
CO
2
control range
0 to 20%
CO
2
variation
± 0.15% (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
O
2
control range
1 to 18%, 22 to 80%
O
2
variation
± 0.2% (ambient temperature; 25
o
C, setting; 37
o
C, 5%, no load)
Chamber humidity
95 ± 5% R.H.
Maximum heat emission
740 kJ/h
Noise level
24 db (A scale)
Usable environment condition
Temperature; 5
o
C to 35
o
C, Humidity; equal or less than 80% R.H.
(The designed performance may not be obtained when
the ambient temperature is equal or less than 15
o
C)
Note: The unit with CE mark complies with EC directives 89/336/EEC, 93/68/EEC and 73/23/EEC.
* It is based on our measuring method.