CALIBRATION
Temperature calibration
1. Press the CAL key for approximately 5 seconds.
2. The third digit of the temperature indicator flashes, and the CO
2
density indicator goes out.
3. Set the present correct temperature with the key and key, then press the ENT key.
4. The unit automatically reverts to the display mode.
[Example]
If the displayed chamber temperature is 37.0
o
C (set value) and the actual temperature is 36.8
o
C.
1. Press the CAL key for about 5 seconds.
2. The “3” on the temperature indicator flashes, and the CO
2
density indicator goes out.
3. Adjust the set value to the actual value of 36.8
o
C with the key and key, then press ENT key.
4. The unit automatically reverts to the display mode.
Note:
It is important to accurately measure the temperature inside the unit when performing temperature
calibration. Particularly, the temperature gauge used must have an accuracy of 0.5 Class or better.
The temperature must be measured at several points.
The temperature setting must not change by more than ±1.0
o
C during calibration. If it exceeds this, an
error tone is emitted, the input data is ignored, and the unit reverts to the display mode. Consequently, if
it is necessary to change the temperature by more than 1.0
o
C, perform calibration in several stages over
a period of time.