Take out the sterilization objects when steam has been purged
completely from the chamber.
12.Take out the sterilization objects.
Set the power switch to OFF at the end of daily use.
13.Set the power switch to OFF.
When taking out the sterilization objects, wear heat-resistant
leather gloves for prevention of burn.
To suspend the heat-retention, press the button. The heat-
retention is suspended. In the case of the melting/heat-
retention cycle, the cycle/program is set at this time
automatically to the 1-1 program of the liquid sterilization cycle.
If the button is not pressed, the heat-retention continues for
72 hours.
For simple heat-retention, change the melting time to 0 in
the melting/heat-retention cycle and press the button.
CAUTION
When taking out the sterilization objects
(objects to be melted) during heat-retention.
For continuous use
When using the product consecutively, make sure on the digital display
I that the temperature inside the chamber is below 60°C.
To check the current temperature inside the chamber in standby
status,press the and buttons simultaneously. (See page 15.)
If the cover is closed when the temperature inside the chamber is
raised due to continuous operation, etc., the air inside the chamber
will rapidly expand, making it difficult to open/close the cover.
Wait until the temperature lowers below 60
°C.
Memo
Drain off the heating water with the cover being opened 2 hours
later or longer after the completion of operation.
14.Drain off the heating water.
1) Open the cover.
2) Open the drain valve.
3) After draining water, close the drain valve without fail.
Repeated use of the same heating water may corrode the
chamber and clog the piping.
Advisory note
2 hours later or longer after
the completion of operation
Drain hose
When the temperature inside
the chamber is 59
CAUTION
22