The temperature inside the chamber is raised hot. When taking
out the sterilization objects, wear heat-resistant leather gloves for
prevention of scald.
CAUTION
To stop the operation
To open the cover immediately
• Lower the pressure and temperature inside the chamber
forcibly in the procedure as follows:
1) Press the button while pressing the button.
2) To stop the exhaust, press the button again.
3) Check the following points before opening the cover.
• Pressure is lowered to 0MPa.
• COVER LOCK lamp goes out.
When the COVER LOCK lamp is lit, the temperature or pressure
inside the chamber rises to make it impossible to pull the
movable handle toward you.
If liquid boils over in liquid sterilization, clean the inside of
the chamber.
Inside of the chamber :
See page 25.
1) Press the button.
•The product is in standby status.
2) Check the following points before opening the cover.
• Pressure is lowered to 0MPa.
• COVER LOCK lamp goes out.
For timer operation
When the COVER LOCK lamp is lit, the temperature or pressure
inside the chamber rises to make it impossible to pull the movable
handle toward you.sterilization objects, wear heat-resistant leather
gloves for prevention of scald.
The operation can be programmed up to at 23:59 of 7 days later from
the current date and time.
1) Press the button.
The lamp above the timer button lights up
and the timer lamp on the digital
display II lights up, and the current date
and time flicker.
2) Press the and buttons to set the operation starting day
and press the button.
• The figures change quickly when the or button is kept
pressed.
• The time on the digital display I flickers.
Lamp goes out.
Beep