-43-
Electrical Adjustments
Group/
Item
Item Name Function Initial Range Note
.. .. - ..
49 Warning Log_50 - 0-32767
50 Warning Log Reset - 0-10
Group
250
FAN Control
0 FAN1 MIN ADJUST (DAC)
DAC Output for Fan
Adjust the tolerance of DAC and Fan
Volage.
* Lamp mode is forced Eco
7 0 - 255
1 FAN1 MAX ADJUST (DAC) 224 0 - 255
2 FAN2 MIN ADJUST (DAC) 13 0 - 255
3 FAN2 MAX ADJUST (DAC) 211 0 - 255
4 FAN3 MIN ADJUST (DAC) 13 0 - 255
5 FAN3 MAX ADJUST (DAC) 224 0 - 255
Group
251
Filter detection setting
0 Filter Warning Calibration 0 0-1
1 Filter Warning Offset Optio-0 1731 0000-9999
2 Filter Warning Offset Optio-1 2686 0000-9999
3 Filter Warning Offset Optio-2 3456 0000-9999
4 Filter Warning Offset Optio-3 4139 0000-9999
5 Filter Warning Offset Optio-4 4728 0000-9999
6 Filter Warning Offset Optio-5 5221 0000-9999
7 Filter Warning Offset Optio-6 5607 0000-9999
8 Filter Warning Offset Optio-7 5899 0000-9999
9 Not Used
10 Not Used
11 Not Used
12 Not Used
13 Filter Warning Times 3 0-10
14 Fan-Add Filter Level 1 1000 0000-9999
15 Fan-Add Filer Level 2 2000 0000-9999
16 Filter Warning Calibration Range 1 0-10
17 Filter Warning Calibration Time 2 0-10
18 Filter Warning Calibration Error Log 255 0-255
19 Filter Warning High Land Calibration Error Log 255 0-255
20 Fliter Warning Calibration Temp 260 -
21
Filter Warning Inclination Rpm (to 26˚C)
60 0000-9999
22
Filter Warning Inclination Rpm (over 26˚C)
80 0000-9999
23
Filter Warning Add Offset Rpm
- -
Group
252
Fan Option
0 Fan Max SWITCH 0: Normal, 1: Max (13.5V) 0 0 - 1
1 SAFETY SWITCH For test purpose 0 0 - 6
2 FAN MANUAL SWITCH 0: Auto, 1: Manual 0 0 - 1
3 FAN1 MANUAL VOLTAGE
Fan Voltage (unit : 0.1V)
Effective only when Fan Maual
switch is 1
100 0 - 145
4 FAN2 MANUAL VOLTAGE 100 0 - 145
5 FAN3 MANUAL VOLTAGE 100 0 - 145
6 FAN4 MANUAL VOLTAGE 100 0-255
7 FAN5 MANUAL VOLTAGE 100 0-255
8 FAN6-8 MANUAL DUTY - 0-511
9 All Fan MaxMin Control 0 0 - 6
10 Press Data
Group
253
Fan Tem Error Setting (Memorized)
Normal Ceiling
0 Temp A Warning (High)
Temp. A to judge the Temp Error at
High (Room)
46 46 30-100
1 Temp B Warning (High)
Temp. B to judge the Temp Error at
High (Panel)
62 62 30-100
2 Temp C Warning (High)
Temp. C to judge the Temp Error at
High (Lamp)
62 62 30-100
3 Temp B-A Warning (High)
Temp. B-A to judge the Temp Error at
High (Clogging Det.)
100 100 0-100
4 Temp C-A Warning (High)
Temp. C-A to judge the Temp Error at
High (Clogging Det.)
100 100 0-100
5 Temp A Warning (Normal)
Temp. A to judge the Temp Error at
Normal (Room)
46 46 30-100
6 Temp B Warning (Normal)
Temp. B to judge the Temp Error at
Normal (Panel)
62 62 30-100
7 Temp C Warning (Normal)
Temp. C to judge the Temp Error at
Normal (Lamp)
60 60 30-100
8 Temp B-A Warning (Normal)
Temp. B-A to judge the Temp Error at
Normal (Clogging Det.)
100 100 0-100
9 Temp C-A Warning (Normal)
Temp. C-A to judge the Temp Error at
Normal (Clogging Det.)
100 100 0-100