Calibration/Adjustment
Available Features
Calibration/adjustment can only be performed when
there is no load on the balance,
the balance is tared,
the internal signal is stable.
If these conditions are not met, an error message is displayed. Otherwise, the weight
required for calibration/adjustment is displayed (see “Accessories” for calibration weights).
You can use any of the following weight units to calibrate/adjust:
g, kg*, lb (menu code 1. 4. x)
You can block calibration/adjustment of the balance:
Select menu code 1. 5. 3)
Example
Step Key (or instruction) Display/Printout
1. Tare the balance t
0.0 g
2. Begin calibration t >2 sec.
+ 1000.0
Calibration weight is displayed
without weight unit (here: 1000 g)
3. Place the indicated calibration
1000.0
weight on the balance
After calibration, the calibration
+ 1000.0 g
weight is displayed with wt. unit
4. Remove the calibration weight
0.0 g
* = not on models with a readability of 0.1 mg