18
ISO/GLP-compliant Printout
Features
You can have the parameters pertaining
to the ambient weighing conditions
printed before (GLP header) and after
(GLP footer) the values of a weighing
series.
These parameters include:
GLP header:
Date
Time at beginning of measurement
Balance manufacturer
Balance model
Balance serial number
Software version number
Identification number of the current
sampling operation
GLP footer:
Date
Time at end of measurement
Field for operator signature
! The record can only be output to
a Sartorius data printer YDP20-0CE.
Settings
G Set the following menu codes
(see “Configuring the Balance”):
GLP-compliant record: menu code 5 5 2
ISO/GLP-compliant record after calibra-
tion/adjustment only:
menu code 7 2 2 or ISO/GLP-compliant
record always on: menu code 7 2 3
Line format for printout: With data ID
codes – 22 characters: menu code 7 1 2
Printout date/time:
– 24-h format: menu code 7 3 1
– 12-h format: menu code 7 3 2
! No ISO/GLP-compliant record is output
if any of the following settings are
configured:
menu codes 6 1 3, 6 1 4 (automatic
printout) and 7 1 1
Function Keys
Press p to output header and first
measured value.
> Header is output the first time p is
pressed
To output header and reference data
automatically with an application pro-
gram active: Press v
End application program:
End application program and output
GLP footer: Press c