12
Weighing in Percent
Purpose
This application program allows you to
obtain weight readouts in percent which are
in proportion to a reference weight.
Preparation
G Configure the Weighing in Percent applica-
tion in the operating menu:
See “Configuring the Balance.”
Menu code: 2. 1. 5
$ Reference percentage:
Code 3. 3. 1 5 %
Code 3. 3. 2 10 % (factory setting)
Code 3. 3. 3 20 %
Code 3. 3. 4 50 %
Code 3. 3. 5 100 %
$ Storage parameter
(display accuracy for counting)
Code 3. 4. 1 Standard resolution: With
stability (factory setting)
Code 3. 4. 2 With 10 times higher
stability than standard
See also “Configuring the Balance”
Example
Determine an unknown percentage: store the weight on the balance as a reference percentage
Menu: Application program: Weighing in Percent (menu code 2. 1. 5)
Menu: Reference percentage: 100 % (menu code 3. 3. 5)
Step Key (or instruction) Display/Printout
1. Tare the balance t 0.0 g
2. Display the reference percentage: v >2 sec. ref100
3. Place the reference weight for 100 % + 222.5 g
on the balance (here: 222.5 g)
4. Start the application; if the print v + 100.00 %
format is set to include data
ID codes the following is printed Wxx% + 222.500 g
5. Place an unknown weight + 144.94 %
on the balance (here: 322.5 g)
6. Display weight v + 322.5 g
7. Display percentage v + 144.94 %
8. Unload the balance 0.00 %
9. Delete the reference percentage c
10. Repeat the procedure starting from step 5, if desired.