7
Selecting the Line Voltage (Mains Voltage)
Use the following original AC adapters for selecting
the line voltage:
AC adapter TNG8, order no. 6971951 (universal)
or
AC adapter TNG8, order no. 6971952 (for the U.K.)
Use the switch to toggle between 230 V and 115 V
Leveling the Balance/Scale
(only for models BL210S, BL120S, BL60S, BL150S,
BL1500S, GC503, GC103, GM1502, GM2202,
GM1202)
Level the balance/scale any time you set it up in
a new location. Use only the 2 front feet of the
balance/scale for leveling.
§ Turn the 2 rear feet until they are in position
(only on models BL1500S and GM1502, GM2202
and GM1202)
§ Turn the 2 front feet as shown here in the illustra-
tion until the air bubble is centered in the level
indicator
> In most cases, this will require several adjustment
steps
Anti-theft Locking Device
To protect against theft, use the mounting lug on
the rear panel of the balance/scale.
Secure the balance/scale at the place of installation,
for example with a chain or a lock