Page #18 background imageLoading...
Page #18 background image
18 Entris
®
BCE Operating Instructions
System Settings
7.3 Parameter List
7.3.1 “SETUP”/”BALANCE” Menu
Parameter Setting values Explanation
AMBIENT V.STABLE Sets the ambient conditions to “very stable”: Activates a fast change in the weight values in
the event of a load change with a high output rate.
Recommended for the following work environment:
Very stable table near the wall
Closed and calm room
STABLE* Sets the ambient conditions to “stable”. Recommended for the following work environment:
Stable table
Slight movement in the room
Slight draft
UNSTABL. Sets the ambient conditions to “unstable”: Activates the delayed change in weight values with
a reduced output rate. Recommended for the following work environment:
Simple office desk
Room with moving machinery or personnel
Slight air movement
V.UNSTBL. Sets the ambient conditions to “very unstable”: Activates a significantly delayed change in the
weight values and long wait for stability with a further reduction in the output rate.
Recommended for the following work environment:
Noticeable and slow floor vibrations
Noticeable building vibrations
Weighed goods moved
Very strong air movements
APP FILT. FINAL.RD.* Activates a filter that enables a fast change in the display for very fast load changes.
Display changes with minimal load changes (in the digit range) occur more slowly.
FILLING Activates a filter that enables a very fast change in the display with minimal load changes,
e.g. when filling containers.
STABIL. V. ACC. Sets the stability to “very accurate”.
ACC.* Sets the stability to “accurate”.
FAST Sets the stability to “fast”.
ZER./TAR. W/O STB. If the button is pressed: The function of the [Zero] or [Tare] button is executed immediately.
W/ STAB.* The function of the [Zero] or [Tare] button is only executed after stability is achieved.
AUTOZER. ON* Activates automatic zeroing. The display is automatically set to zero in case of a deviation of
0 less than (X).
OFF Deactivates automatic zeroing. Zeroing must be triggered with the [Zero] button.
* Factory setting

Table of Contents