Entris
®
BCE Operating Instructions 39
Technical Data
15.10.2 Models BCE2202 | BCE1202
Model
BCE2202-1x (x =
only CCN, OIN) |
BCE2202i-1x
BCE2202-1x (x =
only S, SAR, SJP,
SKR)
BCE1202-1x (x =
only CCN, OIN) |
BCE1202i-1x
BCE1202-1x (x =
only S, SAR, SJP,
SKR)
Unit Value Value Value Value
Readability | Scale interval (d) mg 10 10 10 10
Maximum capacity (Max) g 2200 2200 1200 1200
Repeatability
At 5% load, typical value mg 5 5 5 5
At approx. the maximum load, typical value mg 10 10 10 10
Linearity deviation
Limits ± mg 20 20 20 20
Typical value ± mg 6 6 6 6
Sensitivity drift between +10 °C and +30 °C ± ppm/K 2 3.5 2 3.5
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for i-1x models):
Temperature change K 2 2
Time interval h 6 6
Only for models with approval:
Accuracy class II II
Type BC_EE BC_EE
Verification scale interval (e) mg 100 100
Minimum load (Min) mg 500 500
Minimum initial weighing according to USP (United States
Pharmacopeia), Chap. 41
Optimum minimum initial weighing g 82 82 8.2 8.2
Typical measurement time s
<1.0 <1.0 <1.0 <1.0
Typical stabilization time s
<0.9 <0.9 <0.9 <0.9
Recommended Calibration Weight
External calibrated test weight g 2000 2000 1000 1000
Accuracy class in accordance with OIML R111-1 F1 F1 F1 F1
Weighing pan size mm 182 x 182 182 x 182 182 x 182 182 x 182
Net weight, approx. kg 4.3 | 6.2 4.3 4.3 | 6.2 4.3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius BCE2202 Series and is the answer not in the manual?

Sartorius BCE2202 Series Specifications

General IconGeneral
BrandSartorius
ModelBCE2202 Series
CategoryScales
LanguageEnglish

Related product manuals