Entris
®
BCE Operating Instructions 27
Operation
7.3.22 “LANGUAG.” Menu
Parameter Setting values Explanation
Language ENGLISH,
DEUTSCH,
FRANC.
ITAL., ESPANOL,
PORTUG.,
РYССKИИ,
POLSKI
Defines the menu language.
* Factory setting
8 Operation
8.1 Switching the Device On and Off
Requirements
The device is connected to the power supply.
Procedure
t
CAUTION Pointed or sharp-edged objects may damage
the operating display!
t Only touch the operating display with your fingertips.
t To switch the device on: Press the [On/Off] button.
t To switch the device off: Press and hold the [On/Off]
button.
8.2 Waiting for the Warm-up Time
In order to provide accurate weighing results, the device
must have reached the necessary operating temperature.
The operating temperature is reached after the warm-up time.
The warm-up time starts when the device is switched on.
Procedure
t Switch on the device.
t Wait until the operating temperature is reached (warm-up
time see Chapter “15.7 Warm-up Time”, page 37).
If a weighing process is carried out on a conformity-
assessed device during the warm-up time:
The weight value is marked as invalid.
8.3 Leveling the Device with a Level Indicator
Unevenness at the installation site of the device may result in
incorrect weighing results. Leveling balances out unevenness at
the installation site.
Procedure
1
t Unscrew the rear leveling
feet, until they touch the
setup surface.
y The support foot (1) is not in
contact with the setup
surface.
t To level the device: Bring the
air bubble of the level
indicator into the middle of
the circular marking. To do so,
turn the leveling feet to the
left or right.
t Check that the device is standing on 4 leveling feet and
the support foot (1) is not in contact with the setup
surface.
8.4 Overview of Calibration and Adjustment
During calibration, a calibration weight is used to determine
the deviation of the displayed value from the actual value.
The subsequent adjustment eliminates this deviation.
We recommend regular calibration and adjustment:
Daily, every time the device is switched on
After every leveling
After changing the ambient conditions (temperature,
humidity, or air pressure)
After setting the device up at a new installation site
8.5 Calibrating and Adjusting Device with
isoCAL Function (Only Model i-1x)
Triggers for the automatic start of the isoCAL function are:
The ambient temperature has changed since the last
adjustment process.
The interval time was exceeded (interval time, see Chapter
“15.7 Warm-up Time”, page 37).
If this relates to a conformity-assessed model:
The device has been disconnected from the power
supply since the last adjustment.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius BCE2202 Series and is the answer not in the manual?

Sartorius BCE2202 Series Specifications

General IconGeneral
BrandSartorius
ModelBCE2202 Series
CategoryScales
LanguageEnglish

Related product manuals