Press k to select menu item 11-4-4 (maximum load).
Press ) again to open this menu item; then enter the maximum capacity (for
example 300.0 kg) (key sequence: k k k k ) k ) k )
p ) k). Enter “300” (it is not necessary to enter the decimal point and
the last “0”).
To store the new setting, press and hold the ) key (> 2 sec).
Note: If “Multiple-range scale” or “Multi-interval scale” was selected during A/D
converter configuration rather than “Single-range scale”, then the range limits
must be entered under menu line 11-5 (multiple-range) or 11-6 (multi-interval).
Each of these menu lines has 3 items to choose from, for setting range limits
1, 2 and 3.
Store the configuration data: Select menu line 11-10 (key sequence: ( (
k k k )).
The default setting is 11-10-2, “Do not store configuration”.
Press k to select menu item 11-10-1 (Store configuration).
To store the new settings, press and hold the ) key (> 2 sec).
Display_Serv_1144.eps
Display_Mxl30.eps
Display_Serv_11102.eps
Display_Serv_11101.eps