EasyManua.ls Logo

Sartorius CISL2U - Replacing the A;D Converter; Quick-test of the A;D Converter

Sartorius CISL2U
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
73
Replacing the A/D Converter
Before replacing the A/D converter, try to read out the data from the A/D
converter’s controller using the service software; if this is successful, you can
load the adjustment data in the controller of the new A/D converter once it is
installed. In many cases, this precludes the need to adjust the weighing system.
Because the model designation and serial number are not stored in the A/D
converter’s controller, you do not need to enter this data again after replacing
the A/D converter.
When connecting the weighing platform (load cell(s)) to the A/D converter,
it is important to install a ferrite ring over the signal lines leading from the
platform, to ensure electromagnetic compatibility.
Quick-test of the A/D Converter
Pin Assignment Chart
No. Signal designation Meaning Voltage
1 BR_POS Bridge supply voltage (+) 4.5V ± 0.18V
2 SENSE_POS Sense (+)
Bridge supply voltage
3 OUT_POS Measuring voltage positive
4 OUT_NEG Measuring voltage negative
5 SENSE_NEG Sense (-)
Bridge supply voltage
6 BR_NEG Bridge supply voltage (-) –4.5V ± 0.23V
adu_001.jpg
Combics_18an.eps

Table of Contents