94
Service Specification CAPP/CAPS4U-.....-LU
Example for the order number of a Combics full range scale: CAPS4U-5000KK-RR (Type Designation see page 91-92)
Applications Material | Number of Weighing Platform size Resolution
Design load cells Capacity (Ibs) inches | order code
CAP S 4 U - 5000 RR - LU
just_combics.xls
Reproducibility
Off-center load
Eccentricity
Span Linearity TCS iso CAL
Model
Weighing
capacity
Readability
Tes t
weight
Perm is s ible
tolerance
s
Test
weight
Per miss ible
tolerance
(±)
Class
Adjustm .
weight
Tes t
weight
Perm is s ible
tolerance
(±)
Tareweight Testweight
Pe rm is s ible
tolerance
(±)
ppm
/K
K
CAPP4U-1000..-LU
CAPS4U-1000..-LU
1000 lbs 0,02 lbs 500 lbs 0,2 lbs 500 lbs 0,2 lbs M1 500 lbs 1500 lbs 1 lbs
---
(500)
lbs
300/700/
1100/1500
lbs 0,2 lbs 10 n
CAPP4U-2500..-LU
CAPS4U-2500..-LU
2500 lbs 0,05 lbs 1250 lbs 0,5 lbs 1250 lbs 0,5 lbs M1 2500 lbs 3 lb s 0,002 lbs
---
(1)
lbs 0,7/1,5/2,2/3 lbs 0,5 lbs 10 n
CAPP4U-5000..-LU
CAPS4U-5000..-LU
5000 lbs 0,1 lbs 2500 lbs 1 lbs 2500 lbs 1 lbs M1 5000 lbs 3 lbs 0,002 lbs
---
(1)
lbs 0,7/1,5/2,2/3 lbs 1 lbs 10 n
CAPP4U-10000..-LU
CAPS4U-10000..-LU
10000 lbs 0,2 lbs 5000 lbs 2 lbs 5000 lbs 2 lbs M1 10000 lbs 3 lbs 0,002 lbs
---
(1)
lbs 0,7/1,5/2,2/3 lbs 2 lbs 10 n