63
Note: If a weighing platform has more than one load cell, multiply the nominal capacity
accordingly. Example: The weighing platform consists of 4 load cells, each of which
has a capacity of 30 kg. In this case, the nominal capacity (= maximum load)
equals 4 x 30 kg, or 120 kg.
To store the setting, press and hold the ) key (> 2 sec).
Select menu item1-19-2 (key sequence: k, )).
Enter the resolution (lowest scale interval d) (e.g., 0.01 kg). This value has to
match that stored under menu item 11-4-1 (or 11-5-1 or 11-6-1).
To store the setting, press and hold the ) key (> 2 sec).
Open menu item 1-19-3 (key sequence: k )).
Enter the sensitivity of the load cell in mV/V (usually listed in the load cell
specification sheets). If a weighing platform consists of more than one load
cell, enter the sensitivity of load cell 1.
Example: The sensitivity of the load cell is 1.9440 mV/V.
To store the setting, press and hold the ) key (> 2 sec).
Note: If a weighing platform consists of more than one load cell, enter the sensitivity
values for the other cells (up to 4) under menu items 1-19-4 (for load cell 2)
through 1-19-6 (for load cell 4) and confirm each value (k and ) keys).
If the platform has less than four load cells, enter “0.000” under the unused menu
items.
display_schrittw_001Ib.eps
Display_WZE19440.eps