87
Replacing the Load Cells Foot
– To replace the load cells foot, move the retainer spring to the left hand and right
hand and forth with a screwdriver to push it out.
Color Codes of the Wiring for Weighing Platforms, Models CAPP../CAPS...
Platform size Pin Assignment Chart for the Indicator (see Page 71) 
(1)
or Pin Assignment Chart for the Junction Box (see Page 84) 
(2)
in inches   No.: 1 2 3 4 5 6
12.6 x 9.5 
(1)
Blue Green White Red Gray Black
15.8 x 11.8 
(1)
Blue Green White Red Gray Black
19.7 x 15.8 
(1)
Blue Green White Red Gray Black
24 x 18 
(1 or 2)
Blue Green White Red Gray Black
24 x 24 
(1 or 2)
Blue Green White Red Gray Black
30 x 24 
(1 or 2)
Blue Green White Red Gray Black
30 x 30 
(1 or 2)
Blue Green White Red Gray Black
Platform size Pin Assignment Chart for the Junction Box (see Page 84)
in inches   No.: 1 2 3 4 5 6
(Wiring Bridge          )
36 x 36  Green --- White Red  --- Black
48 x 48  Green --- White Red  --- Black
60 x 48  Green --- White Red  --- Black
60 x 60  Green --- White Red  --- Black
lastz_2.jpg