Operating Instructions Combics Indicators  115
Error Messages
Errors are divided into the following:
–    Dynamic errors are displayed for the duration of the error with an error code (e.g. inf 01).
–    Temporary errors are displayed for 2 seconds (e.g. inf 07) 
–    Fatal errors are displayed continuously (e.g. err101, a reset is required to clear them).
Display  Cause   Solution
No display  No power present   Check power supply 
segments
-------   Key has no function in this status
Flashing k  Battery defective or time changed  Set time
H  Weighing range exceeded  Unload the scale
l or Err 54  Weighing pan is not in place   Position the weighing pan
Err 101 - 104    Key is stuck    Release key or key pressed when switching on the device 
Contact your local Sartorius Service Center 
Err 320  Operating program memory faulty  Contact your local Sartorius Service Center
Err 335    Verified weighing platform not compatible with the   Connect a compatible weighing platform 
connected terminal
Err 340  New EEPROM loaded (Service)    Turn the scale off and then on again. If the error code 
Err340 is still displayed, please contact your local Sartorius 
Service Center
Err 341  RAM has lost data; battery is dead  Leave the scale connected to power for at least 10 hrs.
Err 343    Loss of data in the memory area for transaction numbers   Contact your local Sartorius Service Center 
External alibi memory
Inf 01  Data output not compatible with output format  Set output format correctly
inf 02    Adjustment condition was not met  Do not carry out adjustment until after 0 display 
e.g. not tared or weighing pan loaded  Unload scale, tare using the ) key
inf 03    Adjustment could not be completed within a certain time  Allow to warm up again and repeat the adjustment process
Inf 06  Built-in calibration weight defective  Contact your local Sartorius Service Center
inf 07    Function not allowed in scales verified  
for use in legal metrology  Contact your local Sartorius Service Center
inf 08    The load on the scale is too heavy to zero the readout  Check whether “Tare/zero at power on” (1.12) is set
Inf 09  Taring is not possible when the scale gross weight is < zero  Zero the scale
inf 10    Tare key is blocked when there is data in the tare memory    The data stored for the application program must be 
deleted before taring (Combics 2 only).
inf 18  Preload is too light
Inf 19  Preload is too heavy
inf 29    Minimum load not reached  Reduce min. load 
    (under Application, menu item 3.6)
Inf 30    BPI ID (BPI byte) not deleted in current weighing platform   Reset weighing parameters to factory settings for current  
  (COM1 is fixed on XBPI data communication)  weighing 
platform
Inf 31    Interface handshake interrupted (XOFF, CTS)  Send XON, unblock CTS
inf 71    Cannot store the current weight value  None 
(e.g. control limits too low or too high)
inf 72    Cannot store the current weight value  None 
(e.g. transaction counter maximum reached)
inf 73    Data cannot be written or read  Contact your local Sartorius Service Center
inf 74    Function is blocked (e.g. menu is locked, device is already   None  
configured to another interface)
NO WP  No weighing platform connected  Connect a weighing platform