Calibration, Adjustment
  Purpose   Perform calibration to determine the difference between the value displayed and 
the actual weight on the platform. Calibration does not entail making any changes 
within the weighing equipment.
During adjustment, the difference between the measured value displayed and the 
true weight of a sample is corrected, or is reduced to an allowable level within 
maximum permissible error limits.
    Configuration for Use in Legal Metrology
 
 
Configuration of the weighing instrument for use in legal metrology is set by a 
switch. The switch is located on the back of the weighing platform and covered by 
a protective cap. 
Using a verified scale in legal metrology in the EU:
The Type-Approval Certificate for verified scales is only valid for non-automatic 
weighing instruments. For automatic operation with or without additional, 
integrated equipment, please follow the applicable national regulations for the 
installation location.
  Externally connected IS scales:  Before use in legal metrology, the scale should be calibrated via the internal 
calibration equipment at the installation location: see the “Internal Calibration" 
section in this chapter.
 
3
  The temperature range (°C) listed on the ID label should not be exceeded during 
operation. 
  For Servicing:  External calibration for verified scales of accuracy class l
–   External calibration is blocked in legal metrology (switch cover is sealed)
–   External calibration is only possible by removing the seal. If the seal is broken, 
the validity of verification will become void and you must have your scale  
re-verified.
Using a verified scale in legal metrology with internal calibration equipment:
t   Before use in legal metrology, the “internal calibration" function should be 
carried out at the installation location:
Open the menu access switch
The menu access switch is located on the back of the indicator right next to the 
weighing platform connection.
 
  t  Remove the cap.
t   Slide the switch to the left (= “open" position, for use in legal metrology).
  Operating Instructions Combics Indicators  47
  Operation 
Features 
Which of the following features are available depends on the 
connected weighing platform. These features are 
configured in the Setup menu:
–  External calibration/adjustment blocked in verified 
weighing instruments
–  External calibration/adjustment with the default weight 
value or standard weight (not available on verified 
instruments): setup / wp-1menu   Menu item 1.9. 
“Calibration, Adjustment."
–  Specify the weight for external calibration/adjustment:  
setup / wp-1menu   Menu item 1.18. “Entering 
Calibration Weight."
–  Internal adjustment for IS weighing platforms  
(configure under: COM1 or unicom / WP2), Combics 2
–  Block the J key to prevent use of the two functions 
described above: setup / wp-1menu   Menu item 1.9. 
“Calibration, Adjustment."
–  Calibrate first; then adjust automatically or manually (not 
for verified weighing instruments): setup / wp-1menu   
Menu item 1.10. “Calibration/Adjustment Sequence."
–  Flashing n symbol as adjustment prompt. If more than 
one weighing platform is connected, the platform number 
is also displayed: setup / wp-1menu   Menu item 1.15. 
“Calibration Prompt."
–  Block external calibration/adjustment: setup / wp-1menu   
Menu item 1.16. “External Adjustment."