EasyManuals Logo

Sartorius combics 1 Operating Instructions

Sartorius combics 1
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
Net Total Formulation R (Combics 2)
With this application, you can weigh in different components up to a defi ned total.
Each component is saved in the net-total memory (Applic.3 menu).
Features Weigh in up to 999 components in series
Net total formulation cannot be combined with level 1 and 2 applications
(Applic.1, Applic.2).
Current component number displayed in the number line (indicating the
component to be added)
Toggle the display from “component mode” to “additive mode” by pressing the
w key
Component mode: Display the weight of the component currently on the
platform (for 1 second after it is saved; then the platform is tared)
Additive mode: Display the weight of all components on the platform (after
it is saved, the net weight of the last component added is displayed briefl y)
Toggle to a second weighing platform while weighing in
Activate Info mode via the I key
Automatic component printout when it is saved.
Setting: Applic./Applic.3/NET TOT menu item 3.17.
Printout If the 3.17.2 menu item is selected, the entire component record is printed.
If the 3.17.3 menu item is selected, the following items are generated only once for
the rst component if they have been confi gured:
Blank line, dash line, date, time, ID1 through ID6, header lines 1 and 2. For
subsequent components, each “component” item (“Comp xx”) is followed by a
blank line.
Automatic taring of container weight.
Setting: Applic. / aut.tare menu item 3.7
Restore factory default settings
Setting: Applic. / Reset menu item 9.1
Preparation t Open the APPLIC. / APPLIC.3 /NET TOT menu.
Available parameter settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 digit 3.6.1*
2 digit 2 digits 3.6.2
5 digit 5 digits 3.6.3
10 digt. 10 digits 3.6.4
20 digt. 20 digits 3.6.5
50 digt. 50 digits 3.6.6
100 dig. 100 digits 3.6.7
200 dig. 200 digits 3.6.8
500 dig. 500 digits 3.6.9
1000 dI. 1000 digits 3.6.10
PRT.SAV. Individual/Component printout when saved 3.17.
OFF Automatic printing off 3.17.1
EACH.TIM. Print the entire standard print
confi guration every time the O key
is pressed 3.17.2*
ONCE Print the entire standard print
confi guration once with the O keY 3.17.3
Operating Instructions Combics Indicators 89
Operation

Table of Contents

Other manuals for Sartorius combics 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius combics 1 and is the answer not in the manual?

Sartorius combics 1 Specifications

General IconGeneral
BrandSartorius
Modelcombics 1
CategoryMonitor
LanguageEnglish

Related product manuals