Installation
3
Warning when using pre-wired RS-232 connecting cables: RS-232 cables purchased
from other manufacturers often have pin assignments that are incompatible with
Sartorius products. Be sure to check the pin assignments against the chart in this
manual before connecting the cable, and disconnect any lines identiļ¬ed differently
from those speciļ¬ed by Sartorius.
3
Connect only Sartorius accessories and options, as these are optimally designed for
use with your device. Therefore, do not use any proprietary solutions. The operator
shall be solely responsible for installation and testing of any modiļ¬cations to
Sartorius equipment, including connection of cables or equipment not supplied by
Sartorius. Information on operational quality (in line with norms pertaining to
immunity) is available on request.
t If you have any problems with your device, contact your local Sartorius ofļ¬ce,
dealer or service center.
IP Protection Rating
IP Rating: ā All models are rated to IP44 (IP65 as an accessory).
ā āIP65" models are rated to IP65.
ā The IP65/IP69K protection rating is ensured only if the rubber gasket is installed
and all connections are fastened securely (including the caps on unused
sockets). Weighing platforms must be installed and tested by a certiļ¬ed
technician.
ā If you install an interface port or battery connection after setting up your
indicator, keep the protective cap in a safe place for future use. The cap protects
the interface connector from vapors, moisture and dust or dirt.
Use in Legal Metrology
ā When the indicator is connected to a weighing platform and this equipment is
to be veriļ¬ed, ensure that the applicable regulations regarding veriļ¬cation are
observed.
ā When connecting Sartorius weighing platforms, observe the permitted weighing
range as listed in the āGuide to Veriļ¬cation of Weighing Instruments" and the
Declaration of Conformity.
ā A sticker with the āSartorius" logo was afļ¬xed to the indicator as a control seal
following veriļ¬cation. This seal will be irreparably damaged if you attempt to
remove it. This will nullify the veriļ¬cation's validity. In this case, re-veriļ¬cation
would be required in compliance with all relevant national regulations and laws.
Operating Instructions Combics Indicators 5
Warnings and Safety Precautions