In the Applic./Applic.3/CLASS./Param.2/OUTP.ACT menu, menu item 4.7,
you can make the following settings for the control outputs:
switched off
always on
on at stability
The “SET” output normally changes its voltage level when the current weight
exceeds the minimum load. Alternatively, you can assign the “Ready for use”
function to this port.
Setting: Applic./Applic.3/CLASS./PARAM.2/CTRL.SET menu item 4.3.
Example 2: There should be three classes.
Configuration: The “Classification" application is selected, and printout has been
set up.
O t Enter the class limits using the O key.
.
1 1 0 t Enter the upper limit for Class 1 using the keypad (in this example, 110 g)
.
+
O t Save the upper limit for Class 1.
1 3 0 t Enter the upper limit for Class 1 using the keypad (in this example, 110 g).
O t Save the upper limit for Class 1.
t Place the sample on the weighing platform.