Loading...
t Press J to start an external adjustment.
y The display “C.EXT.D" appears briefly.
In the example, the altitude and latitude of the installation location are being entered.
y The display “Altitud" appears briefly.
y The altitude at the place of installation is displayed in meters above sea level - here, the
altitude for “Germany (Zone D).”
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y The display “Latitud" appears briefly.
y The geographical latitude of the place of installation is shown in degrees north or
degrees south - here the latitude setting for “Germany (Zone D).”
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y You are prompted to place the required weight on the platform (e.g.: 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described in the
chapter entitled “Operation” under “Calibration and Adjustment.”
30
Configuring Weighing Platforms
C.EXT.D
C.EXT.DEF.
Altitud
C.EXT.DEF.
Latitud
C.EXT.DEF.

Table of Contents