EasyManua.ls Logo

Sartorius Combics 3 CAIS3 - Gravity Input and Calibration Completion; Entering Gravity for Adjustment; Finalizing External Calibration;Adjustment

Sartorius Combics 3 CAIS3
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
31
Configuring Weighing Platforms
If gravity is being entered instead of altitude and latitude, then “Gravity" is displayed
for a brief time after “
CAL."
The entered value appears in m/s
2
, here for the “Germany (Zone D)" setting.
t Press J to confirm the displayed value or press ( to cancel the adjustment.
y You are prompted to place the required weight on the platform (e.g.: 5.0 kg).
The subsequent steps for completing the calibration/adjustment are described in the
chapter entitled “Operation” under “Calibration and Adjustment."
t Slide the menu access switch to the right (= “closed" position) and reattach the cap.
y The display goes out and the device restarts. Then weighing mode is active.
If adjustment is carried out using a verifiable configuration data record, the lines
for display of metrological data (lines 1 and 2) show the data valid for use in legal
metrology, if the menu access switch is closed. See also the chapter “Operation",
“Configuration for Use in Legal Metrology."