14
Leveling the Balance
Purpose:
– To compensate for unevenness at the place
of installation
Leveling Balances with a Weighing Capacity
up to 10 kg
Only the 2 front feet are adjusted to level the
balance.
§ Retract the two rear feet (only on models with
a rectangular weighing pan).
§ Turn the 2 front feet as shown in the diagram
until the air bubble is centered within the circle of
the level indicator.
> In most cases this will require several adjustment
steps.
§ For weighing heavy samples: Extend the 2 rear
feet until they touch the surface on which the
balance rests (only on models with a rectangular
weighing pan).
Leveling Balances with a Weighing Capacity
over 10 kg
§ Adjust the leveling feet until the air bubble
is centered within the circle on the level indicator.