EasyManuals Logo

Sartorius CP Series User Manual

Sartorius CP Series
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #44 background imageLoading...
Page #44 background image
Weighing in Percent
44
Preparation
Set parameters for the Weighing in
Percent program:
$ Select the application program in the
Setup menu
§ Set the following parameters:
2 Application programs
2 1 Program selection
2 1 5 Weighing in percent*
3 Application parameters
3 5 Storage parameter
3 5 1 ο Standard resolution
(internal resolution)
3 5 2 10+ higher resolution
3 6 Decimal places for calc.
3 6 1 None
3 6 2 ο 1 decimal place
3 6 3 2 decimal places
3 6 4 3 decimal places
ο = Factory setting
* = Not in CPA…-PCE models
Menu code: 2 1 5 *
Display symbol: %
Purpose
This application program allows you to
obtain weight readouts in percent which
are in proportion to a reference weight.
Printout: Weighing in Percent
pRef + 100 % : Reference percentage
Wxx% + 111.6 g: Reference weight net xx%
Prc + 94.7 %: Calculated percentage
WWW.BALANCES.COM