Parameter Settings (Overview)
o Factory setting
User-defined setting
Setup 1 Weighing 1 1 Adapt filter 111 Very stable conditions
112 o Stable conditions
113 Unstable conditions
114 Very unstable conditions
1 2 Application filter 121 o Final readout
122 Filling mode
1 3 Stability range 131 4 digit
The stability symbol is 132 1 digit
displayed when the 133 1 digit
value is stable within 134 o 2 digits
this number of digits. 135 4 digits
136 8 digits*
1 5 Tare function* 151 Without stability
152 o After stability
1 6 Auto zero 161oOn
162 Off
1 7 Weight unit 1 171 Grams (display: o)*
172 o Grams (display: g)
173 Kilograms
1
)
174 Carats
175 Pounds*
176 Ounces*
177 Troy ounces*
178 Hong Kong taels*
179 Singapore taels*
1710 Taiwanese taels*
1711 Grains*
1712 Pennyweights*
1713 Milligrams
2
)
1714 Parts per pound*
1715 Chinese taels*
1716 Mommes*
1717 Austrian carats*
1718 Tola*
1719 Baht*
1720 Mesghal*
1 8 see next page
2 through 9 see following pages
* = Not available in balances verified for use in legal metrology
1
) = Not available for models with a readability < 0,1 mg
2
) = Not available for verified balances of accuracy class K
28
Menu level 1
Menu level 2
Menu level 3
Factory setting
Menu item