Animal Weighing/Averaging
46
Preparation
Set parameters for the
Animal Weighing program:
$ Select the application program
in the Setup menu
§ Set the following parameters:
2 Application programs
2 1 Program selection
2 1 7 Animal weighing*
3 Application parameters
3 7 Animal activity
3 7 1 Calm (2% of the
animal/object)
3 7 2 ο Normal (5% of the
animal/object)
3 7 3 Strong vibration (20%
of the animal/object)
3 8 Start
3 8 1 Manual
3 8 2 ο Automatic
ο = Factory setting
* = Not in CPA…-PCE models
Printout: Animal Weighing
mDef 20 : Number of subweighing operations
x-Net + 401.1 g: Calculated average
Menu code: Code 2 1 7 *
Display symbol: V
Purpose
Use this program to determine the weights
of unstable samples (e.g., live animals)
or to determine weights under unstable
ambient conditions. With this program,
the balance calculates the weight as the
average of a defined number of individual
weighing operations (also referred to as
“subweighing operations").