22
Internal Calibration/Adjustment
The menu code setting 1 9 3
must be selected in the Setup menu.
Inside the balance housing is a
motorized calibration weight which
is applied and removed automatically
for internal calibration.
G Activate calibration: Press (CAL)
> The built-in calibration weight is
applied automatically
> The balance is calibrated
> If “Calibrate, then auto adjust in one
operation" is selected in the Setup
menu, the balance is now adjusted
automatically.
> The internal calibration weight is
removed.
! Important note for calibration/adjust-
ment of the CPA2P-F model: position
the cover before calibration/adjust-
ment
Calibration and Adjustment Sequence
In the Setup menu, you can configure
whether:
Calibration is always followed auto-
matically by adjustment (1 10 1; factory
setting), or
You have the choice of ending the
sequence or starting adjustment after
calibration (1 10 2)
If no difference is determined between
nominal and actual weights, you can
end the calibration routine following
calibration.
Two keys are active at this point:
(CAL) = start adjustment
(CF) = end the sequence