52
The ISO/GLP-compliant printout can contain the following lines:
-------------------- Dotted line
17-Jan-2007 10:15 Date/time (beginning of measurement)
SARTORIUS AG Balance manufacturer
Mod. CPA10001 Balance model
Ser. no. 10105355 Balance serial number
Ver. no. 00-13-47 Software version
ID 2690 923 ID
-------------------- Dotted line
LID Measurement series no.
nRef + 10 pcs Counting: Reference sample quantity
wRef + 21.14 g Counting: Reference weight
Qnt + 235 pcs Counting result
Qnt + 567 pcs Counting result
-------------------- Dotted line
17-Jan-2007 10:20 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line
ISO/GLP-compliant printout for external calibration/adjustment:
-------------------- Dotted line
17-Jan-2007 10:30 Date/time (beginning of measurement)
SARTORIUS AG Balance manufacturer
Mod. CPA10001 Balance model
Ser. no. 10105355 Balance serial number
Ver. no. 00-13-47 Software version
ID 2690 923 ID
-------------------- Dotted line
Cal. Ext. Calibration/adjustment mode
Set + 5000.0 g Calibration weight
Diff. + 0.2 g Difference determined in calibration
Cal. Ext. Complete Confirmation of completed calibration procedure
Diff. + 0.0 g Difference from target following adjustment
-------------------- Dotted line
17-Jan-2007 10:32 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line