EasyManuals Logo

Sartorius CP Series User Manual

Sartorius CP Series
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
52
The ISO/GLP-compliant printout can contain the following lines:
-------------------- Dotted line
17-Jan-2007 10:15 Date/time (beginning of measurement)
SARTORIUS AG Balance manufacturer
Mod. CPA10001 Balance model
Ser. no. 10105355 Balance serial number
Ver. no. 00-13-47 Software version
ID 2690 923 ID
-------------------- Dotted line
LID Measurement series no.
nRef + 10 pcs Counting: Reference sample quantity
wRef + 21.14 g Counting: Reference weight
Qnt + 235 pcs Counting result
Qnt + 567 pcs Counting result
-------------------- Dotted line
17-Jan-2007 10:20 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line
ISO/GLP-compliant printout for external calibration/adjustment:
-------------------- Dotted line
17-Jan-2007 10:30 Date/time (beginning of measurement)
SARTORIUS AG Balance manufacturer
Mod. CPA10001 Balance model
Ser. no. 10105355 Balance serial number
Ver. no. 00-13-47 Software version
ID 2690 923 ID
-------------------- Dotted line
Cal. Ext. Calibration/adjustment mode
Set + 5000.0 g Calibration weight
Diff. + 0.2 g Difference determined in calibration
Cal. Ext. Complete Confirmation of completed calibration procedure
Diff. + 0.0 g Difference from target following adjustment
-------------------- Dotted line
17-Jan-2007 10:32 Date/time (end of measurement)
Name: Field for operator signature
Blank line
-------------------- Dotted line
WWW.BALANCES.COM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius CP Series and is the answer not in the manual?

Sartorius CP Series Specifications

General IconGeneral
BrandSartorius
ModelCP Series
CategoryScales
LanguageEnglish

Related product manuals