106 Operating Instructions Cubis
®
MCA
Technical Data
15.2 Power Supply
15.2.1 Device
Only by Sartorius AC adapter YEPS03-15V0
15.2.2 AC Adapter
Unit Value
Item No. YEPS03-15V0
Primary
AC voltage V 100–240 (± 10%)
Frequency Hz 50–60 (± 5%)
Current consumption, maximum A 1.0
Secondary
DC voltage at 2 A output current V 14.25–15.75
Power, maximum W 30
Short circuit protection: Electronic
Protection class according to IEC 62368-1 I
Pollution level according to IEC 61010-1 2
Overvoltage category according to IEC 606641-1 II
Installation site according to IEC 62368-1,
maximum altitude above sea level
m 5000
Temperature
In operation °C 0–+40
During storage and transport °C -20–+80
Connector for mains connection according
toIEC 60320-1/C14:3-pin
Power supply cable
Power supply cable according to IEC 60320-1/C14:
Country-specific, 3-pin, two-sided plug
Other data: See label on AC adapter
15.2.3 Safety of Electrical Equipment
According to EN 61010-1/IEC 61010-1 Safety requirements for
electrical equipment for measurement, control, and laboratory use –
Part 1: General Requirements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA5203S Series and is the answer not in the manual?

Sartorius Cubis MCA5203S Series Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA5203S Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals