Operating Instructions Cubis
®
MCA 113
Technical Data
15.7.4 Models MCA3203S | MCA2203S | MCA2203P | MCA1203S
MCA3203S MCA2203S MCA2203P MCA1203S
Unit Value Value Value Value
Scale interval (d) mg 1 1 1 | 10 1
Maximum capacity (Max) g 3200 2200 1010 | 2200 1200
Repeatability at 5% load
Standard deviation of the load values,
tolerance
mg 1 0.7 0.7 0.7
Standard deviation of the load values,
typical value
mg 0.5 0.5 0.5 0.5
Repeatability at approx. maximum capacity
Standard deviation of the load values,
tolerance
mg 1 1 1 0.7
Standard deviation of the load values,
typical value
mg 0.6 0.6 0.6 0.6
Linearity deviation
Tolerance mg 5 3 5 2
Typical value mg 2 2 3 1
Deviation when load is off-center,
positions according to OIML R76
Test weight g 1000 1000 1000 500
Tolerance mg 2 2 3 2
Typical value mg 1 1 2 1
Sensitivity drift between +10°C and +30°C ppK 1 1 1 1.5
Tare maximum capacity:
Less than 100% ofmaximum capacity
Accuracy class according
toDirective2014/31/EU
I I I I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 10 10 10 10
Minimum load (Min) according
toDirective2014/31/EU
mg 100 100 100 100
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing mg 820 820 820 820
Typical minimum initial weighing mg 1000 1000 1000 1000
Typical stabilization time s 1 1 1 1
Typical measurement time s 2 1.5 1.5 1.5