116 Operating Instructions Cubis
®
MCA
Technical Data
15.9 isoCAL Function
15.9.1 Models MCA225S | MCA225P | MCA125S | MCA125P | MCA324S | MCA324P |
MCA224S | MCA124S | MCA2203S | MCA2203P | MCA1203S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling
15.9.2 Models MCA524S | MCA524P | MCA5203S | MCA5203P | MCA3203S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 1.5
After a time interval h 6
After successful leveling
15.9.3 Models MCA623S | MCA623P | MCA323S | MCA5202S
Unit Value
isoCAL is triggered by the following criteria:
In the event of a temperature change K 2
After a time interval h 12
After successful leveling
15.10 Data Storage Device
Maximum number of datasets 150,000