DQD 507 Rev. 2016-02-18 Page 2
APPLICABLE REQUIREMENTS
CSA Standards:
CAN/CSA-C22.2 No. 61010-1-12 - Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use, Part 1: General Requirements
UL Standards:
UL Std. No. 61010-1 (3
rd
Edition) - Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use - Part 1: General Requirements
CONDITIONS OF ACCEPTABILITY
(1) Equipment is intended to be used with power supply FSP Group Inc, Model FSP030-DGAN3 or Sinpro
Electronics Co Ltd, Model SPU31-106 or PSFD rated output 15Vdc, 2A. The equipment can also be used
with any other power supply approved for Canada and USA under the 60950-1 or 61010-1 or equivalent
standards and which provides Reinforced insulation between mains and secondary circuits. It must be
rated for 3000m and up to 40°C and must provide an output of 15Vdc LPS with a minimum of 1A
(2) Detachable cord set, to be delivered with the product, consisting of minimum NEMA 5-15 Plug, 3 x
18AWG cable and C13 Appliance coupler. For countries outside Canada and the USA the cable must be
equivalent and must fulfill the national requirements for detachable cord sets.