Example: Determining animal weight with automatic start and 20 subweighing operations
(measurements)
Parameter settings: Applic.: AnimalW. (menu code 2. 7.)
Step Key (or instruction) Display/Data output
1. Place animal weighing bowl
22.6 g
on the balance/scale
2. Tare the balance/scale
w 0.0 g
3. Change the number of subweighing
operations:
b ref 30
4. Set number of measurements: Repeatedly: b
In increments of 1 (1, 2, 3, etc. to 100) Press briefly
In increments of 10 (10, 20, etc. to 100)
b press ref 20
and hold
5. Confirm number of measurements
u + 0.0 g
*
and start automatic animal weighing
The number of measurements
remains stored in battery-backed
memory until the setting is changed
6. Place first animal in bowl.
888
The balance/scale delays the start 20
of measurements until the difference 19
between 2 measurements ....
meets the criterion 1
7. Read off the result + 410.1 g
G
*
The result is displayed with the
ā€œ*ā€ symbol (= calculated value)
mDef 20
and remains displayed until x-Net + 410.1 g
the sample (animal) is removed
from the load plate (bowl)
8. Unload the balance/scale +
0.0 g
*
9. Weigh next animal (if des.)
Next weighing series begins automatically