Calibration/Span Adjustment
Calibrate the Scale
Step Key (or instruction) Display
a. Switch on the scale (On/Off)
b. Zero the scale (Zero/Left)
c. Select calibration/linearization (Tare/Right) > 2 sec
mode
d. Select calibration (Fn/Up)
repeatedly
e. Confirm/start calibration. (Print/Enter)
After the zero point is stored,
the required calibration weight
is displayed
f. Place the required weight
on the scale
g. If the weight is applied within (Print/Enter)
the defined time limit and
tolerance, the “OK” symbol
is displayed.
Press “(Print/Enter)” to confirm
and save calibration.
(To exit calibration press
(Zero/Left) > 2 sec)