Miras 2
Miras 2
31
Error Codes
Error codes are shown on the main display. “Err“ codes are shown continuously;
“Inf.“ messages are shown for 2 seconds,
after which the program returns automatically to the weighing mode.
Problem Cause Solution
Nothing appears No power available Check the power supply
or some symbols The AC adapter is not plugged in Plug in the AC adapter
incomplete Battery is drained Replace battery; charge battery using
on the display LCD cable is loose external charger
Plug in the cable again or contact
customer service
oL The load exceeds the scale capacity Unload the scale
uL Weighing pan is not in place Place the weighing pan on the pan
Something is touching the weighing pan Move the object that is touching the
weighing pan
d5PErr Display overflow:
Value cannot be shown on the display Reduce the load on the scale
CALErr Calibration parameter not met; e.g.: Calibrate only when zero is displayed
scale not zeroed Press (zero) to tare the scale
scale is loaded Unload the scale
scale setting not allowed
to the customer
APPErr Weight is too high or there is no sample Increase the weight on the scale
on the scale with application in use
PrtErr Data interface for printing is blocked Contact customer service
5y5Err APP board cannot retrieve data Contact customer service
from AD board
EPErr EEPROM defective Contact customer service
FnErr Function not allowed in scales Contact customer service
verified for use in legal metrology
5b1Err SBI command error Contact customer service
Err 08 Zero is not possible if weight value Zero the scale after removing
is more than 10% of FSD some of the load
Err 09 Tare is not possible when the gross Zero the scale
weight is a minus value
FEErr Fatal error; cause unknown Contact customer service
Err 54
Err 55 No platform is connected Connect a platform or sensor
Max. Weighing capacity The scale was switched on without Place the weighing pan on the scale
is less than indicated the weighing pan in place and press “On/Off”
under “calibrate”
The weight readout The scale was not calibrated/ Calibrate/adjust the scale
is obviously wrong adjusted before weighing Zero the scale
Scale not zeroed