EasyManua.ls Logo

Sartorius M-Power AZ214 - Weighing in Percent

Sartorius M-Power AZ214
29 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Weighing in Percent
Purpose
This application program allows you to
obtain weight readouts in percent which are
in proportion to a reference weight.
Preparation
Configure the Weighing in Percent applica-
tion in the operating menu:
See “Configuring the Balance.”
Menu code: 2. 1. 5
$ Reference percentage:
Code 3. 3. 1 5 %
Code 3. 3. 2 10 % (factory setting)
Code 3. 3. 3 20 %
Code 3. 3. 4 50 %
Code 3. 3. 5 100 %
$ Storage parameter
(display accuracy for counting)
Code 3. 4. 1 Standard resolution: With
stability (factory setting)
Code 3. 4. 2 With 10 times higher
stability than standard
See also “Configuring the Balance”
Example
Determine an unknown percentage: store the weight on the balance as a reference percentage
Menu: Application program: Weighing in Percent (menu code 2. 1. 5)
Menu: Reference percentage: 100 % (menu code 3. 3. 5)
Step Key (or instruction) Display/Printout
1. Tare the balance t 0.0 g
2. Display the reference percentage: v >2 sec. ref100
3. Place the reference weight for 100 % + 222.5 g
on the balance (here: 222.5 g)
4. Start the application; if the print v + 100.00 %
format is set to include data
ID codes the following is printed Wxx% + 222.500 g
5. Place an unknown weight + 144.94 %
on the balance (here: 322.5 g)
6. Display weight v + 322.5 g
7. Display percentage v + 144.94 %
8. Unload the balance 0.00 %
9. Delete the reference percentage c
10. Repeat the procedure starting from step 5, if desired.
www.balances.com is your Authorized Sartorius Dealer
www.scaleman.com is your Authorized Sartorius Dealer

Related product manuals