Model AZ214 AZ124 AZ64
Weighing capacity g 210 120 60
Readability mg 0.1 0.1 0.1
Tare range (subtractive) g 210 120 60
Repeatability <±mg 0.2 0.2 0.2
Linearity <±mg 0.3 0.3 0.3
Operating temperature range +10…+30 °C (50°F to 86°F)
Allowable ambient
operating temperature +5…+40°C (41°F to 104°F)
Sensitivity drift
within +10…+30 °C (50°F–86°F) >±/K 2 · 10
–6
2 ·10
–6
2 ·10
–6
Response time (average) s 3 3 3
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update (depends on the
filter level selected) s 0.2–0.4 0.2–0.4 0.2–0.4
External calibration weight g 200 (E2) 100 (E2) 50 (E2)
(of at least accuracy class) lb 0.4 0.2 0.1
Net weight, approx. kg 3.2 3.2 3.2
Pan size mm 90 d 90 d 90 d
Weighing chamber height mm 200 200 200
Dimensions (WxDxH) mm 200+270+299
AC power source/
power requirements AC adapter 230 V or 115 V, +15% to –20%
Frequency 48–60 Hz
AC power source,
direct current V 10–20
Power consumption (average) W 1 1 1
Hours of operation with the
YRB08Z rechargeable battery pack h 20 20 20