Example: setting the following drying parameters
Heating temperature: 130°C
Start of analysis: without stability after the cover is closed
End of analysis: after 10 minutes
Display mode for result: moisture
Step Key (or instruction) Display
1. Turn on the analyzer e Self-test runs
2. Select PRG: drying program
parameters
3. Confirm PRG 105
°
C
(the previously set temperature
is displayed; in this example, 105°C)
4. Set the heating temperature
repeatedly 130
°
C
(in this example: 130°C)
5. Confirm heating temperature
0.0
min
(the previously set analysis time
is displayed; in this example, 0.0 min)
6. Set the parameter for the
repeatedly 10
min
end of analysis; in this example,
10 minutes)
7. Confirm the “end of analysis”
parameter
8. Select the result display mode
or
%M
(in this example, moisture)