27
Operation
Example: Analysis with Specified Drying Time
The specified drying time in the example is 10 minutes.
Heating temperature:  130°C
Start of analysis:  Without stability after the cover is closed
End of analysis:  After 10 minutes
Display mode:  Moisture
Step  Key (or instruction)  Display
1.  Switch on the moisture    e  Self-test runs
 analyzer   
2.  Set the drying parameters 
  (see “Setting the Drying 
  Parameters” in the chapter 
  entitled “Configuration”)
3.  Open the sample chamber and 
  place an unused sample pan  
  on the pan support 
 
4.  Tare the sample pan:    or    
  select TAR   as needed 
  ...and confirm 
5.  Distribute approx. 2 g sample    
  evenly on the sample pan 
6.  Close the sample chamber   
 
  The printout header is printed: 
  see next page