29
Adjusting the Analyzer
  Heating Element Adjustment
The procedure for adjusting the heating 
element is described in the instructions 
supplied with the YTM15MA temperature 
adjustment set.
  Weighing System Adjustment
  To adjust the weighing system, perform 
calibration and adjustment as described 
in the following.
 Purpose
  Calibration is the determination of the 
difference between the weight readout 
and the true weight (mass) of a sample. 
Calibration does not entail making any 
changes within the weighing system.
  Adjustment is the correction of the  
difference between the measured value 
displayed and the true weight (mass)  
of a sample, or the reduction of the  
difference to a level within specified 
 permissible error limits.
 Features
  Calibration is performed externally with 
the following weight value:
–  MA35: 30 g; see “Accessories,” order no. 
YSS43-02
  You can have calibration and adjust-
ment results documented as a ISO/GLP- 
compliant printout (see the page after 
next for an example).