26
RC-F-32.TIF
RC-F-29.TIF
RC-F-32.TIF
RC-F-30.TIF
Installing the Draft Shield
– Remove the guide bushing (107) for the new draft shield (111) from the
centering ring (108).
– Install the new draft shield (111) and adjust it.
– Fasten the new draft shield (111) in this position through the base plate
using the screws (113).
– Route the connecting cables (E) of the draft shield drive through the retain-
ers on the enclosure frame (208/211) and insert the plug back into the
socket of the digital PCB (204).
– Install the guide bushing  (107), align it according to the pan adapter
(430) and tighten it on the centering ring (108).
Important Note:
Use the guide bushing (107) from the old draft shield when installing a new
draft shield on balances of model series RC 250 S up to serial number
10602262.  The reason:  these types of balances have a thicker pan
adapter (430).
– Close the balance housing (see page 24).
– Now replace the shield disk (106), shield ring (105), weighing pan
(103) and glass cylinder (102) of the weighing chamber draft shield).
– Finally, place the draft shield cover(s) (101 / 104) on the draft shield
(111). Now the balance is ready to operate.