Operation
46
(
  1.) Zero the scale.
J  2.)  Start calibration (e.g., when 
adjustment prompt flashes 
WP).
     c.ext.def appears for two 
seconds.
     You are prompted to place 
the required weight on the 
platform (e.g., 10,000 g).
  3.)  Position the calibration/
adjustment weight on the 
weighing platform.
     The difference between the 
measured value and the true 
weight of the sample will be 
displayed with plus/minus 
signs.
Ext.     calibration A printout will be generated 
Nom.  +      10000 g if the process is cancelled 
Diff. +          1 g    using the ( key.
 
J  4.)  Activate calibration/adjust-
ment (press the ( key to 
cancel).
     The adjustment weight is 
displayed once adjustment is 
complete.
--------------------  A GMP-compliant printout is
24.02.2011     10:15  generated.
TypSIWR
Ser.no.     12345678
Vers.    1.0103.11.2
BVers.      01-26-02
-------------------- 
Ext.     calibration 
Nom.  +      10000 g 
Diff. +          1 g
Ext.      adjustment 
Diff. +          0 g
--------------------
24.02.2011     10:15
Name:
-------------------
Example:
External calibration and manual adjustment with default weights (weighing parameters: factory settings)