Signum 1-2-3
Neutral Measurement Application
Z
nM
Neutral Measurement Application Z
nM
56
Neutral Measurement
Application Z nM
With this application you, can use
your weighing platform to measure the
length, surface and volume of parts.
The o symbol is displayed as the
weight unit.
Characteristics
Save the reference weight “wRef” from
the weighing platform
Signum
®
3 only:
Enter the reference weight “wRef”
through the keypad
Signum
®
3 only:
Enter the factor for calculation “nRef”
using the keypad
Enter reference sample weight using a
barcode scanner
Measure with two weighing platforms
Activate Info mode via the I key
Toggle the display between
measurement and weight via the
w key
Define the level of accuracy (display
resolution) applied when a calculated
reference weight is saved
Automatic taring of container weight.
Configured in Setup under:
Appl: Z nm: 3.7.
(autotare first weigh)
Signum
®
3:
Appl: a.tare: 3.7.
Automatic initialization when the
scale is switched on. The indicator is
initialized with the most recently used
calculation factor “nRef” and reference
weight “wRef.” Configured in Setup
under:
Appl: Z nm: 3.8.
(Start App. with last values)
Signum
®
3:
Appl: a.start: 3.8.
Exit application, delete parameters:
The value of the reference sample
weight in the reference memory remains
active until the application is changed
or the value is overwritten or deleted
via the c key. The reference sample
weight also remains saved after the
scale is turned off.
Signum
®
3 only:
You can assign different functions to
the c key for deleting applications.
When you clear applications, you can
delete either the data stored for all
applications or just the data stored for
the active application. Configured in
Setup under:
Appl: sel.cf: 3.24.
(Sel. CF function key c in
applications)
Tare function:
1) If you store a tare (weight value)
by pressing the ) key, you can later
enter a value manually. The value you
enter is added to the stored tare value.
Setting: menu code 3.25.1 (factory
setting)
2) A value entered manually overwrites
a stored tare value (weight value). If you
enter a value manually, a tare value
(weight value) stored later overwrites
the manually entered value. Setting:
menu code 3.25.2
Configured in Setup under:
Signum
®
2:
Appl
: Z nM: 3.25. ;
Signum
®
3:
Appl
: tare. f: 3.25.
Restore factory default settings.
Configured in Setup under:
Appl: Z nM: 9.9.1.
Signum
®
3:
Appl: def.app: 9.1.
In order to calculate the length, surface
or volume of a given sample, the
average weight of a reference quantity
of the sample must be known (in the
example below, the reference is 1 meter
of electrical cable). There are three ways
to enter the reference weight in the
program:
Calculation:
Place the reference quantity (defined
by the calculation factor) on the
connected weighing platform and
calculate the reference sample weight
by pressing the O key.
Place any amount of the sample
material on the connected weighing
platform, enter the calculation factor
through the keypad, and press the
O key to calculate the reference
sample weight.
How the reference weight is calculated
depends on the application setting for
resolution. The value is either rounded
off in accordance with the display
resolution, or saved with 10-fold
or 100-fold resolution, or with the
maximum internal resolution of the
weighing platform.
Keypad input: Enter the reference
weight (i.e., the weight of one meter of
electrical cable) using the keypad and
press
O to save it.
Use a barcode scanner
The initial application values remain
active until deleted by pressing the
c key or until overwritten by a new
value. They remain saved after the scale
is switched off.
Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select the Setup menu: Press the
k key repeatedly until appl is
displayed
§ Confirm appl: Press the ) key
§ Select the Neutral Measurement
application:
Press the k key repeatedly and
confirm with the ) key
Neutral Measurement Application Parameters
3.6. Minimum load for initialization
3.6.1* 1 digit
3.6.2 2 digits
3.6.3 5 digits
3.6.4 10 digits
3.6.5 20 digits
3.6.6 50 digits
3.6.7 100 digits
3.6.8 200 digits
3.6.9 500 digits
3. 6.10 1000 digits
3.9. Reference value calculation
resolution
3.9.1* Display accuracy
3.9.2 Display accuracy
+1 decimal place
3.9.3 Display accuracy
+2 decimal places
3.9.4 Internal resolution
3.10. Decimal places in displayed result
3.10.1* none
3.10.2 1 decimal place
3.10.3 2 decimal places
3.10.4 3 decimal places
3.11. Parameter for saving weight values
3.11.1* With stability
3.11.2 With increased stability
3.13. Reference weighing instrument
3.13.1* Off
3.13.2 Weighing platform
WP 1
3.13.3 Weighing platform
WP 2
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times.